Search

Prouduct

Capsules With HT Tuffryn® Membrane

HT 멤브레인 캡슐 Consumable > Filter > Capsule Filter
제품설명
  • 정확한 충전, 균일 한 흐름 및 우수한 처리량을 얻습니다. 내장 된 에어 벤트는 갇힌 기포를 제거
  • 제품 및 시스템 무결성을 보장하기 위해 무결성 테스트 가능
  • 감마선 조사에 의한 살균은 EtO 잔류 물에 의한 잠재적 오염을 제거
연락처
경기본사 : 031-790-1907 (내선1)
대전지사 : 042-824-5167
 *Capsules with HT Tuffryn membrane  
P/N Description Pkg
12158 Micro culture capsule with filling bell,
0.2/0.2μm, 300cm2, gamma-irradiated
1/pkg
12140 Culture capsule, 0.2/0.2μm, 500cm2,
gamma-irradiated
1/pkg
12170 Culture capsule with filling bell, 0.2/0.2μm,
500cm
2, gamma-irradiated
1/pkg
12141 Maxi culture capsule, 0.2/0.2μm, 1290cm2,
gamma-irradiated
1/pkg
     
 *Accessories and replacement parts   
P/N Description Pkg
4227 Filling bell, sterile 5/pkg
*Applications
- 1~10L 범위의 batches 여과에 사용(micro culture capsule) 및
완충액 및 세포배양배양액 10~20L에 사용(culture and maxi culture capsules)
Materials of Construction Filter media: HT Tuffryn membrane(polysulfone)
Housing and filling bell: Polycarbonate
Vent plug and core: Polypropylene
Filter support: Polyester
Sealing materials: Polyurethane
Effective filtration area PN 12158: 300cm2
PN 12140 and 12170: 300cm2
PN 12141: 300cm2
Inlet/Outlet connections PN 12140, 12170 and 12141: Hose barb fittings accept 12.7mm (1/2 in.) ID tubing
PN 12158: Stepped hose barbs 6.4 - 9.5mm
(1/4 - 3/8 in.) diameter
Typical Hold-up volume with 0.7 bar(70kPa, 10psi) air purge
PN 12158: 15mL
PN 12140  and 12170: 25mL
PN 12141: 113mL
Maximum operating temperature 100℃(212℉)
Maximum operating pressure 3.4 bar (340kPa, 50psi) at ambient temperature
Recommended Integrity test minimum
bubble point-Water
2.4 bar (240kPa, 35psi)
Typical water flow rate  L/min at 0.7bar(70kPa, 10psi)
PN 12158: 1.2
PN 12140, and 12170: 2.0
PN 12141: 5.2
Bacterial retention Lot samples retain a minimum of
107 cfu/cm2 of B.
Diminuta per modified ASTM F838, current revision
Endotoxin level < 0.5 EU/mL using Limulus Amoebocyte Lysate
(LAL) test
Biological safety Passes Unitied States Pharmacopeia (USP)
Biological Reacitivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation.
If desired, autoclave once only prior to use at
121- 123℃(250-253℉) for a maximum of 20 min.