디지털 자동 선광계 Analyzer > Polarimeter > Visual & Automatic Circle Polarimeter
제품설명
  • High-performance circle polarimeter
  • Unchanged highest precision
  • Continuous measurement
  • For versatile applications
  • Maker : SCHMIDT+HAENSCH / GERMANY
연락처
경기본사 : 031-790-1907 (내선1)
대전지사 : 042-824-5167

Automatic circle polarimeter for standard applications continuously measuring

Fully electronic circle polarimeter with following features:

  • Accurate measuring of the optical activity of solutions according to ICUMSA, O.I.M.L., Australian Standard K157, European and American Pharmacopeia
  • Direct reading of the rotation angle without mechanical transmission is resulting in guaranteed unmatchable precision over the entire measuring range
  • Recommended for routine analysis
  • Short measuring times, no mechanical abrasion
  • The measurement is carried out constantly without pressing a button so that flow through measurements can easily be performed
  • The customer can program up to 10 special scales (glucose, mono sodium glutamate or other) and 4 methods
  • LED illumination unit, life time guarantee
  • Automatic temperature compensation if using appropriate polarimeter tubes
  • Indication of the optical density
  • Serial to Ethernet converter for data transfer to PC
  • The instrument has a versatile software for test routines and configurations
  • GLP/GMP conform documentation
  • 21 CFR part 11 ready, appropriate software “Aquisys 2008“ optionally available
  • High quality “Made in Germany“, consultation and service at site possible

Checking/Calibration

As the interference filters, which are used in circle polarimeters, suffer from age- and environment depending alteration, it might be necessary to check the wavelength accuracy periodically and to readjust the wavelength in case it’s out of range. This can be done with sucrose solutions or quartz control plates since the optical properties of sucrose solution and quartz are almost identical. Sucrose solutions are not very stable and have to be renewed regularly. Checking a polarimeter by a quartz plate is a much more reliable method.

Quartz control plates are long-term stable examination standards for polarimeter. The sample room of SCHMIDT + HAENSCH polarimeter permits the vertical movement of quartz control plates as well as the free rotation (±360°) around their own axis. Only by a rotation it is possible to check if the plates are mounted without strain and permit an examination of the polarimeter according to processing regulations of international standards like ICUMSA and OIML.
 

Main specifications

Measurement scale °Optical Rotation / °Specific Rotation / °Z International Sugar Scale / % Concentration / freely definable scales
Measuring range ± 360° / ± 259°Z
Resolution 0,01° / 0,05°Z
Precision ±0,01° / ±0,05°Z
Reproducibility 0,01° / 0,05°Z
Sensitivity up to OD 3 / up to OD 5
Response time approx. 4 sec
Wavelength one or two fixed
Measurement mode continuous

Temperature

Temperature measurement If temperature sensor with appropriate tubes is used / T-Cell Peltier tubes
Temperature control sample with appropriate tubes via external water bath / Peltier temperature controlled sample tubes
Temperature control range depends on the water bath / 18 – 40 °C
Resolution depends on the water bath / 0.01 °C
Precision depends on the water bath / ± 0.1 °C
Reproducibility depends on the water bath / 0.05 °C

Additional information

Measuring tubes up to 200 mm length
Ambient temperature 10 – 40 °C
Light source LED
Display Graphics-LCD 16 x 16 characters
Operation Alpha numerical keyboard, 20 buttons inclusive function keys
Interfaces/communication 2 serial RS-232, 1 parallel, USB* und Ethernet*

Standard models

  Polartronic N100
Polartronic N101
Polartronic N202

Conformity

Standards International Pharmacopoea, ASTM, AOAC, DIN, FDA, ICUMSA und weitere
  • VariPol C
  • Small Universal Polarimeter
  • Polartronic M series